Experience in using ALPHA-VRA-30 software for determination of iron and zinc content in organometallic compounds and polymers
https://doi.org/10.26896/1028-6861-2018-84-8-20-24
Abstract
X-ray fluorescence analysis (XRF)based on a technique of direct external standard with the use of one reference sample is traditionally used in the laboratory of microanalysis (INEOS RAS) for determination of metals in composition of novel organometallic compounds and polymers. Preparation of the samples-emitters consists in 150 – 200-fold dilution of the substances with emulsion polystyrene (PS), since the amount of substance for analysis is limited to 10 – 15 mg. With this dilution XRF analysis are carried out in the concentration range of metal from 0.005 to 0.5 %, where the analytical calibration curve is linear. Measurements of analytical lines are carried out on a VRA-30 spectrometer (CARL ZEISS JENA (Germany)), X-ray tube with Rh-anode, 40 kV, 20 mA. A possibility of automatic conduction of X-ray fluorescence analysis of organometallic compounds and polymers synthesized in INEOS RAS on a VRA-30 spectrometer using software ALPHA-VRA-30 is studied. Software package developed at the research and innovation center JSC «FORATEKS» (Ekaterinburg) is intended for automated control of the spectrometer functioning without an operator. The software provides the possibility of deducing analytical equations of multiple regression on the basis of measurements of the intensities of analytical lines from the reference sample, and calculates the content of chemical elements in the analyzed samples-emitters. Advantages of the developed software are demonstrated and the results of XRF obtained using both the developed software and the method of direct external standard are compared. The small difference between the results obtained by those methods (0.7 % (abs)) in determination of iron and zinc in organometallic compounds when the content of iron and zinc range within of 4 – 25 % proves the applicability of the software to studying new compounds. Software ALPHA-VRA-30 provides standardization of measurement conditions and lightens the work of the analyst when analyzing large batches of substances.
About the Authors
V. N. TalanovaRussian Federation
Valeriya N. Talanova
Moscow
O. L. Lependina
Russian Federation
Olga L. Lependina
Moscow
D. Kh. Kitaeva
Russian Federation
Dinara Kh. Kitaeva
Moscow
A. G. Buyanovskaya
Russian Federation
Anastasiya G. Buyanovskaya
Moscow
S. I. Dzvonkovski
Russian Federation
Sergey L. Dzvonkovski
Moscow
References
1. Gel’man N. I. (Ed.). Methods of quantitative elemental microanalysis. — Moscow: Khimiya, 1987. — 293 p. [in Russian].
2. Losev N. F., Smagunova A. N. Fundamentals of X-ray fluorescence analysis. — Moscow: Khimiya, 1982. — 207 p. [in Russian].
3. Bakhtiyarov A. V., Savel’ev S. K. X-ray fluorescence analysis of mineral raw materials. — St. Petersburg: Izd. S.-Peterburg. Univ, 2014. — 132 p. [in Russian].
4. Talanova V. N., Lependina O. L., Buyanovskaya A. G., et al. Sources of errors in nondestructive X-ray fluorescence analysis of small samples diluted with a solid diluent: XRF determination of Mn in cymantrenes / Zavod. Labor. Diagn. Mater. 2017. Vol. 83. N 10. P. 65 – 69 [in Russian].
5. Kuzmina T. G., Troneva M. A., Kononkova N. N., Romashova T. V. Error of sample preparation in pressing emitters for X-ray fluorescence analysis / J. Anal. Chem. 2017. Vol. 72. N 3. P. 272 – 278.
Review
For citations:
Talanova V.N., Lependina O.L., Kitaeva D.Kh., Buyanovskaya A.G., Dzvonkovski S.I. Experience in using ALPHA-VRA-30 software for determination of iron and zinc content in organometallic compounds and polymers. Industrial laboratory. Diagnostics of materials. 2018;84(8):20-24. (In Russ.) https://doi.org/10.26896/1028-6861-2018-84-8-20-24